6
REPAIR KITS
Analog Part# Digital Part#
PM3000 / PM3000D Vac Reg RK6000 RK6000D
PM3000E / PM3000DE Vac Reg RK6000E RK6000DE
PM3000HV / PM3000DHV Vac Reg RK6000HV RK6000DHV
PM3000DEHV Vac Reg N/A RK6000DEHV
PM3100 / PM3100D Vac Reg RK6100 RK6100D
PM3100E / PM3100DE Vac Reg RK6100E RK6100DE
PM3100HV / PM3100DHV Vac Reg RK6100HV RK6100DHV
PM3100EHV / PM3100DEHV Vac Reg RK6100EHV RK6100DEHV
PM3200E Vac Reg RK6200E N/A
PM3500 / PM3500D Vac Reg RK6500 RK6500D
PM3500E / PM3500DE Vac Reg RK6500E RK6500DE
PM3600 / PM3600D Vac Reg RK6600 RK6600D
PM3600E / PM3600DE Vac Reg RK6600E RK6600DE
Not all Repair Kits are listed above, contact Precision Medical, Inc. for
availability.
DISASSEMBLY INSTRUCTIONS
1. Loosen the Set Screw (Item# 11) in Selector Knob.
2. Pull the Control Knob Assembly (Item# 12) away from case. (The
Regulator Module (Item# 10) is threaded onto the Control Knob Assembly.)
3. Remove the screws (Item# 2) from the back of the Regulator.
4. Separate the Case Assembly (Item# 9) by pulling it away from the
Housing Assembly (Item# 1).
5. Remove the Selector Assembly (Item# 6) by pulling it away from the
Housing Assembly (Item# 1).
6. Remove the Gauge Assembly (Item# 3).
ASSEMBLY INSTRUCTIONS
in
reverse order.
NOTE:
aligned and engaged when reassembling.
2. Lubricate all O-rings and cavities with Vacuum grease (part# 1775)
supplied in the Vacuum Regulator Repair Kit.
4.
Prior to returning Vacuum Regulator to service verify accuracy of gauge.