CLEANING / DECONTAMINATION
(As needed)
1. Attach a working Vacuum Regulator with a continuous regulated
mode to a minimum vacuum source of 15 inHg.
2. Mix cold disinfection/sterilization solution according to its

3. Connect tubing as shown in Cleaning Illustration on previous page.
4.
Turn the working Vacuum Regulator on to a continuous regulated mode.
5. Adjust the vacuum to a minimum of 120 mmHg.

set at 100 mmHg.
7. Allow cold disinfection/sterilization solution to pass through and
collect in Suction Canister. Procedure should continue for time
recommended by the manufacturer of the cold disinfection/sterilization
solution for the desired level of disinfection or sterilization.
(if
applicable).
9. Allow remaining cold disinfection/sterilization solution to pass
through and collect in Suction Canister.
10. Set working Vacuum Regulator to its maximum vacuum setting.
11. Thoroughly dry the internal components by drawing maximum
vacuum through the Regulator to be cleaned for at least 30 seconds

NOTE:
If it is not possible to pass cold disinfection/sterilization
solution through the Regulator, then the passageways
are totally blocked and DISASSEMBLY of the Regulator is

MAINTENANCE
Before each use, visually inspect Vacuum Regulator for any sign of
damage, DO NOT USE if damaged.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number,
contact Precision Medical, Inc. All returns must be packaged in sealed
containers to prevent damage. Precision Medical, Inc. will not be responsible
for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy
available on the Internet, www.precisionmedical.com.